Quality standards CE, ISO13485, ISO9001, Design patent certification |
Application Range Adult/Pediatric/Neonatal/Medicine/Surgery/ Operating Room/ICU/CCU |
Display 7” real color TFT screen |
Environment Operating environment: Temperature: 0~40°C Humdidty: 15%~80% Pressure: 86.0kPa~106.0kPa Transport and Storage environment: Temperature: -20~55°C Humdidty: 10%~90% Pressure: 86.0kPa~106.0kPa |
Power Requirements AC: 100~240V,50Hz/60Hz DC: Built-in rechargeable battery Battery: 11.1V 24wh Li-ion battery 2h operating after full charge 5min operating after low battery alarm |
Dimension and Weight Equipment: 175mm*140mm*300mm; 3.3kg Package: 360mm*250mm*300mm; 4.5kg |
ST segment Measurement range: 0~100% Resolution: 1% Accuracy: 70%~100% ±2%, 40%~69% ±3% |
Pluse Rate Measure/alarm range: 20~300bpm Resolution: 1bpm Accuracy: ±3bpm |
Displayed Parameters Time: Battery-backed quartz crystal clock ECG: ECG waveform scale, displayed lead Heart Rate: Derived from ECG or SpO2 NIBP: Pressure (systolic, mean and diastolic) Pulse Oximeter: Pulse Rate, SpO2 waveform, and Oxygen saturation. Respiration Rate: Respiration rate derived from ECG. Trends: HR, RR, NIBP, TEMP and SpO2 Temperatures: one channels |
ECG 5 Leads: RA, LA, LL, RL, V Lead mode: I, II, III, aVR, aVL, aVF, V Gain: x1.25mm/mV, 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV (error≤±10%) Bandwidth: Surgery 1~25Hz Monitor 0.5~40Hz Diagnostic 0.05~130Hz Sensitivity: >200 μV P-P CMRR: ≥112dB Differential Input Impedance: >5M ohm Electrode Offset Potential: ±300mV Polarization Voltage: ≥±500mV Imput Impedance: ≥5M Calibrating Voltage: 1mV ±5% |
RESP Method: impedance between RA-LL Resp Impedance range: 0.3~3Ω Base Impedance range: 200Ω~4000Ω Bandwidth: 0.3~2.5Hz Gain: x0.25, x0.5, x1, x2, x4 Resp rate: Measurement Range Adult: 0~120BrPM Neonatal/Pediatric: 0~150BrPM Resolution: 1BrPM Accuracy: 7~150BrPM (±2BrPM) 0~6BrPM Apena Alarm: 10~40 seconds Neonatal: SYS 40~135mmHg |
NIBP Method: oscillometry Work mode: Manual/Auto/STAT Measure interval in auto mode: 1,2,3,4,5,10,15,30,60,90,120,180,240,480min Measure period in STAT mode: 5min Pulse rate range: 40~240bpm Measure & alarm range: Adult: SYS 40~280mmHg DIA 10~220mmHg MEAN 20~240mmHg Pediatric: SYS 40~220mmHg DIA 10~160mmHg MEAN 20~170mmHg Neonatal: SYS 40~135mmHg DIA 10~100mmHg MEAN 20~110mmHg Static pressure measurement Range: 0~300mmHg Accuracy: ±3mmHg Resolution: 1mmHg Accuracy Max. average error: ±5mmHg Max. standard deviation: 8mmHg Overvoltage protection: Adult 300mmHg Pediatric 240mmHg Neonatal 150mmHg |
TEMP (optional) Channel: 1 Measure & alarm range: 0~50°C Resolution: 0.1°C Presision (no sensor): ±0.1°C (0~50°C) |
Standard accessories · NIBP cuff & tube · ECG cable & electrodes · SpO2 sensor · TEMP probe · Lithium-ion battery · Power cable · Operator’s manual |
Optional accessories · TEMP probe (skin) · Trolley bracket · Hanging bracket · Monitor recorder · Vehicle-mounted power supply |